Tempsens Calibration Centre is an
independent unit of Tempsens instruments (I) Pvt. Ltd.
We are the only private sector company in the country owning
Fixed Point Temperature calibration Laboratory with NABL
Accreditation. The lab has highly stable calibration furnaces,
measuring instruments and accurate master sensors traceable to
National and international Standards. The calibration center
functions as per ISO17025/NABL standards.
Calibration of contact and non contact type sensors can be made
in temperature range of -38°C to 1500°C & 50°C to 1500°C.
Further the laboratory offers both in house and on site
temperature calibration.
Unique Advantages
One of the best Labs in India equipped with various type of
calibration furnaces and measuring instruments.
Fixed point cells calibration
Provides best B.M.C.
NABL Accreditation
Provide on-site calibration
Prompt and Fast Services
Facilities
Fixed Point Calibration
We are first to install fixed point temperature calibration for Triple point of water (Tpw), Gallium (Ga), Tin (Sn), Zinc (Zn) & Aluminum (Al) Cells.
Fixed points are most accurate devices available for defining a temperature scale. Fixed point cells are designed to realize the liquid-solid equilibrium temperatures of certain high purity metal elements, for calibration of thermometers at ITS-90 fixed points.
On site calibration
Our experienced and trained Engineers go to the site to calibrate the instrumentation, usually without the time and expense of removing and reinstalling them. This saves time and money of customer.
Our engineers help customers to provide fast and accurate on-site calibration for contact and non contact temperature sensors in temperature range -25°C to 1200°C & 50°C to 1200° with lowest down time.
High Accuracy Imported Calibrated Master Sensors
Wide range of calibration equipments- set of Dry blocks, Liquid baths along with very accurate temperature sensors.
Trained Staff
Undertaken intensive internal and external training in India and abroad.
Simple Temperature Calibration Software
Dedicated simple and cost effective temperature calibration software which supports Tempsens Calibrators.
Computerized Automatic Calibration Report
Proprietary software is used to generate reports cross checks various fields in the calibration record. When calibration certificates are issued the instruments and the sensors are tested against precise standard parameters
regarding the measuring accuracy.
Calibration Seminars / Training
We conduct training and consultancy for temperature solution to cater needs of customers.
Continual Innovation
Upgradation of system & solution continually
SCOPE OF ACCREDITATION (Permanent Facility)
S.#
Parameter Measured
Range
B.M.C
Instrument Used
1.
Contact Type Temp Sensors (Thermocouples, RTD's, Digital Temp, Indicator, Thermometers)
-38....0 °C
0....140 °C
140....250 °C
250....500 °C
500....1000 °C
1000....1200 °C
1200....1500 °C
±0.07 °C
±0.04 °C
±0.08 °C
±0.14 °C
±1.1 °C
±2.1 °C
±4.2 °C
Using Cropico Digital Thermometer,
PRT(25.5) with Methanol Liquid Bath,HART Bath, Oil Bath
& Calsys 650
Using Cropico Digital
Thermometer & PT/RH-PT T/C with Calsys 1200L & Calsys
1500L Dry Block Bath.
2.
Non-contact Infrared
Pyrometers
50....
250 °C
250.... 400 °C
400....500 °C
500....800 °C
800....1000 °C
1000....1200 °C
1200....1500 °C
±0.40 °C
±0.90 °C
±2.60 °C
±3.50 °C
±4.10 °C
±4.10 °C
±5.80 °C
Using Impac Black Body, Cropico
Digital Thermometer, RTD 7 Standard Pyrometer
Using LAND Black Body & Standard Pyrometer
3.
Contact Type
Temp. Sensor (For Fixed Point Cells)
(0.01°C)
29.7646°C
231.928°C
419.527°C
660.323°C
±0.015°C
±0.029°C
±0.039°C
±0.033°C
±0.036°C
Using TPW Cell with HART Maintenance Apparatus
Using Ga Cell *
Using Sn Cell *
Using Zn Cell *
Using Al Cell *
SCOPE OF ACCREDITATION (On Site Facility)
S.#
Parameter Measured
Range
B.M.C
Instrument Used
1.
Contact Type Temp Sensors (Thermocouples, RTD's, Digital Temp, Indicator,
Thermometers)
-25°C…0°C
0°C…..140°C
140°C…250°C
250°C…500°C
500°C…1000°C
1000°C…1200°C
±0.06°C
±0.07°C
±0.10°C
±0.14°C
±1.42°C
±2.20°C
Using Cropico Digital Thermometer,
PRT(25.5) with Methanol Liquid Bath,HART Bath, Oil Bath
& Calsys 650
Using Cropico Digital Thermometer & PT/RH-PT T/C with
Calsys 1200L & Calsys 1500L Dry Block Bath
2.
Non-contact
Infrared Pyrometers
50....
250 °C
250.... 400 °C
400....500 °C
500....800 °C
800....1000 °C
1000....1200 °C
±0.40 °C
±0.90 °C
±3.30 °C
±3.50 °C
±4.10 °C
±4.10 °C
Using IMPAC Black Body ,Cropico
Digital Thermometer, RTD & Standard Pyrometer
Using Calsys 1200 TBB Black Body , Standard Pyrometer