PRINCIPLES AND APPLICATIONS OF RESISTANCE THERMOMETERS AND THERMISTOR
Resistance thermometers and Thermistor are temperature sensors, which
changes there electrical resistance with temperature. The superior sensitivity
and stability of these devices, in comparison to thermocouples, give them
important advantages in low and intermediate temperature ranges. In addition,
resistive devices often simplify control and readout electronics.
Resistance thermometers are specified primarily for accuracy and stability from
cryogenic levels to the melting points of metals. They are accurate over a wide
temperature range and may be used to sense temperatures over a large area,
and are highly standardized. The standard platinum resistance thermometer is
specified by the ITS-90 (International Temperature Scale of 1990) to interpolate
between fixed points in the range 13.80K (-259.35°C) to 1234.93K (961.78°C).
RESISTANCE THERMOMETERS
Resistance thermometers may be called as RTDs (resistance temperature
detectors), PRT's (platinum resistance thermometers), or SPRT's (standard
platinum resistance thermometers). These thermometers operate on the
principle that, electrical resistance changes in pure metal elements, relative to
temperature.
The traditional sensing element of a resistance thermometer consists of a coil of
small diameter wire wound to a precise resistance value. The most common
material is platinum, although nickel, copper, and nickel-iron alloys compete
with platinum in many applications.
A relatively recent alternative to the wire-wound RTD substitutes a thin film of
platinum or nickel-iron, which is deposited on a ceramic substrate and lasertrimmed
to the desired resistance. Thin film elements attain high resistances
with less metal, thereby lowering cost.
RESISTANCE/TEMPERATURE CHARACTERISTICS
Resistance thermometers exhibit the most linear signal with respect to
temperature of any sensing device.
Small deviations from straight-line response, however, dictate the use of
interpolating polynomials to calculate resistance values between fixed
temperature points.
The resistance/temperature characteristic for standard platinum resistance
thermometers, as defined by the ITS-90, is a complex set of equations. Platinum
generally follows the modified Callendar-Van Dusen equation over the range
-200 to 850°C (-328 to 1562°F):
TEMPERATURE COEFFICIENT OF RESISTANCE
Temperature Coefficient of Resistance (TCR) has many definitions. For
resistance thermometers, TCR is normally defined as the average resistance
change per °C over the range 0 to 100°C, divided by R °C: 0
In one sense, TCR expresses the sensitivity of the resistive wire used in the
element, as it defines the average temperature change of a hypothetical 1
thermometer.
But specifies normally use TCR to distinguish between different
resistance/temperature curves of the same element material, such as the three
curves platinum. Because all of these curves see widespread use, platinum
TCR's must be properly specified to maintain compatibility between
thermometers and instruments.

There are four primary curves specified for platinum:
1. 0.003926 /°C: Standard platinum resistance thermometers are the only
PRT's that can achieve this TCR. They must have high purity platinum wire
(99.999% or better) wound in a strain-free configuration.
The stresses introduced in manufacturing, lower the TCR of ordinary industrial
models.
Several manufacturers offer industrial platinum thermometers with nominal TCR
of 0.00392; TCR's around 0.003923 are achieved regularly.
2. 0.003911 /°C: This TCR is sometimes called the “U.S. Industrial Standard.”
It is lower than laboratory standards as the typical construction of high
temperature ceramic elements impose strain on platinum wire.
3. 0.00385 /°C:
DIN 43760, IEC 751, and other national and international specifications
mandate this TCR.
4. 0.00375 /°C: Elements with 0.00375 TCR, intended for low-cost
applications.
There are few inherent advantages in specifying any particular TCR over another.
Laboratory systems traditionally use reference standards with the highest-grade
platinum, but industrial specify may aim instead for the greatest degree of
standardization. In this case, 0.00385 TCR will be compatible with the greatest
number of manufacturers.
COMPARISON OF ELEMENT TYPES
Platinum, with its wide temperature range and stability, has become the
preferred element material for resistance thermometers.
Furthermore, advances in element construction have narrowed the price
difference between platinum and base metal thermometers. Nevertheless,
nickel, copper, and nickel-iron do have benefits for many applications and should
be considered. The primary advantages of the four element types are compared
in Table.

Table : Comparison of resistance thermometer element type
EFFECTS OF LEAD WIRE RESISTANCE
Because an RTD is a resistance type sensor, any resistance in the extension
wires between RTD and the control instrument will add to the readings. In some
cases, one can compensate for this extra resistance with adjustments at the
instrument. However, this only compensates when the leads are at a constant
temperature since variations in ambient temperature alter copper lead wire
resistance.
Below mentioned Table shows resistance values of common copper lead wire
according to sizes.

Lead wire error can be significant, especially with small diameter leads or with
low sensitivity elements.
In a 2-lead bridge, the lead wire resistance adds directly to readings. If leads are
short enough or if sensitivity is high enough, the offset may be acceptable.
When long extension runs are required between the sensor and instrument, or
sensitivity is low, the specified should consider a 3-lead system. All resistance
thermometers with copper elements must have three leads to offset their low
sensitivity.
3-lead systems represent a practical compromise between accuracy over distance
and the cost of extra leads. Although it is well suited to most industrial areas,
electrical noise and contact resistance at the junction points may affect them.
4-lead circuits provide the same resistance compensation as 3-lead systems,
but also relieve problems with unmatched leads, contact resistances, and
thermal EMF's.
The one disadvantage of using 4-wire ohms is that we need one more extension
wire than the 3-wire bridge.
This is a small price to pay if we are at all concerned with the accuracy of the
temperature measurement.
POTENTIAL SOURCES OF ERROR WITH RESISTANCE
THERMOMETERS
Resistance thermometer systems are susceptible to three types of errors: The
inherent tolerances built into the thermometers, gradients between the
thermometer and the medium to be sensed, and errors introduced along the
path between the sensor and readout or control instrument. Some sources of
error are electrical; others result from the mechanical construction of the
thermometer.
1. INTERCHANGEABILITY/CONFORMITY:
Conformity specifies the
amount of resistance a thermometer is allowed to deviate from a standard
curve (such as the curve produced by the Callendar-Van Dusen equation).
A tolerance at the reference temperature, usually 0°C, and a tolerance on the
slope or TCR. Figure 5 shows that a resistance thermometer conforms most
closely to its curve at the reference temperature, while the resistance fans out
above and below this reference.
For example, IEC 751, Class B, requires calibration within 0.12 (0.3°C) at 0°C, but
allows TCR to deviate from nominal 0.00385 by ±0.000012 /°C. Thus, tolerance
spreads to 0.8°C at 100°C, 1.3°C at 200°C, and on up to 3.8°C at 700°C.
Interchangeability between two thermometers is no more than twice the value
of there Conformity. Commercial platinum resistance thermometer elements are
available with extremely tight tolerances, to within 0.01 (0.026°C) in some
cases. When interchangeability is an overriding consideration, the specified may
consider other means to achieve it. For example, manufacturers may alter their
calibration procedures to fix the reference temperature and tightest tolerance at
a point other than 0°C. Or if the difference between two thermometers is more
important than absolute temperature, matched pairs measured to agree within
a certain tolerance may is less expensive than calibrating each thermometer
within a small range of nominal.

It is important to note that conformity/interchangeability specifications only
denote the relative accuracy of two otherwise identical thermometers mounted
side by side in the same environment. They do not include errors acting equally
upon both thermometers.
2. SENSITIVITY:
The resistance change per degree change in temperature is
a function of base resistance and TCR (Temperature Coefficient of Resistance).
Although a thermometer with higher sensitivity is not necessarily more
accurate, a larger signal simplifies output electronics and is less susceptible to
lead wire effects and electrical noise. In addition, a larger resistance produces
the same voltage output with less measuring current, which helps to limit selfheating
of the thermometer element.
3. INSULATION RESISTANCE:
If the sensing element and leads are not
completely insulated from the case, a shunting effect occurs in which the case
becomes a parallel resistor and lowers apparent readings. In most industrial
thermometers, with specified insulation Resistances in the 100-megohm ranges, error approaches zero. The
manufacturer must take care to seal water-absorbing materials.
The shunting effect decreases with low-resistance elements, which accounts for the use of 25.5 PRT's in laboratory measurements.
4. SELF-HEATING:
A resistance thermometer is a passive resistance sensor;
it requires a measuring current to produce a useful signal. Because this
measuring current heats the element wire above the true ambient temperature,
errors will result unless the extra heat is dissipated.
Self-heating is most often expressed in mW/°C, which is the power in mill watts
2 (1000 I R) required to raise the thermometers internal temperature by 1°C. The
higher the mW/°C figure, the lower the self-heating. As an example, assume a 5
mA measuring current is driven through a 100 platinum RTD at 100°C. Selfheating
is specified as 50 mW/°C in water moving at 3 ft/sec. The amount of
heat generated is:
2 1000 mW x (0.005 A) x (138.5 ) = 3.5 mW
The self-heating error is:
(3.5 mW) / (50 mW/°C) = 0.07°C
The generated heat increases with higher sensor element resistance (when a
constant current measurement device is used), or with increasing measuring
current.
The resulting error is inversely proportional to the ability of the thermometer to
shed extra heat; which, in turn, depends on thermometer materials,
construction, and environment.
The worst self-heating occurs when a high resistance is packed into a small
body. Thin film elements, with little surface area to dissipate heat, are an
example. Self-heating also depends on the medium in which the thermometer is
immersed. Error in still air may be over 100 times greater than in moving water.
5. TIME CONSTANT:
A time constant indicates the responsiveness of a
resistance thermometer to temperature change. A common expression is the
time it takes a thermometer to reflect 63.2% of a step temperature change in
moving water. Response speed depends on the mass of the thermometer and
the rate at which heat transfers from the outer surface to the sensing element.
A rapid time constant reduces errors in a system subject to rapid temperature
changes.
6. REPEATABILITY:
The degree of accord between two successive readings
with a thermometer is its repeatability.
Loss of repeatability results from permanent or temporary changes to the
resistance characteristics of the element and may be caused by exposing the
thermometer to temperatures at or beyond the endpoints of its specified range.
A repeatability test cycles the thermometer between low and high
temperatures; any changes to R are noted. A typical repeatability rating for an 0°C
industrial platinum resistance thermometers is ±0.1°C.
7. STABILITY:
Stability is long-term drift in thermometer readings.
A typical specification would limit drift to 0.1°C per year for rated operation.
Normal services at points well within the temperature rating typically cause
much less drift. Drift is a consequence of the element material, with platinum
being the most stable; encapsulating materials, which could contaminate the
element; and mechanical stress placed on the element by expansion of winding
bobbins or other supporting structures.
8. SHOCK AND VIBRATION:
Mechanical shock and vibration can alter
thermometer readings or cause complete failure.
In fact, stability and ruggedness are somewhat exclusive. A laboratory
thermometer designed for maximum stability contains an unsupported element,
which is far too fragile for industrial use.
The elements of most industrial resistance thermometers are fully supported by
a bobbin or packing material, and therefore stand up quite well to extreme
environments.
More likely to suffer are lead wire transition points, which should be properly
immobilized. A typical RTD will meet a specification allowing shock of 100 G's of
8 milliseconds duration and vibration of 10 to 2000 Hz at 20 G's.
9. PACKAGING AND THERMAL TRANSFER:
Sheaths and other structures
surrounding resistive elements should maximize heat transfer from the sensed
medium, minimize heat transfer from ambient which can alter readings, and
provide necessary protection of the elements.
Proper materials and construction can dramatically improve reading accuracy.
One strategy practicable only with wire-wound resistance thermometers versus
thermistors, thermocouples, and solid-state devices is temperature averaging.
An element may be wound to average temperature over lengths of up to 100
feet.
RESISTANCE THERMOMETER TYPES
Examples of commonly available resistance thermometer types, with an
emphasis on the design features which take advantage of the benefits listed
above, and which avoid the sources of error, are presented in the following
paragraphs.
1. Standard Platinum Resistance Thermometers for Laboratory Use
NIST specifies the standard platinum resistance thermometer (SPRT) as the
standard interpolating instrument used to define temperatures from -259.35 to
961.78°C. According to the ITS-90, a standard platinum Resistance
thermometer must meet one of the following criteria:

Figure shows a classic “bird cage” strain-free element, in which insulating disks
supports the element wire. Figure shows a less expensive helical wind. The
highest accuracy primary standards encapsulate the element within a glass or
quartz sheath, although stainless steel or Inconel may be substituted for
secondary, or transfer, standards. The element area must be hermetically sealed
to exclude oxidizing agents and helium filled if used at cryogenic ranges.

The unsupported coils of a strain-free element are quite susceptible to damage
from shock and vibration. Even a slight tap on a table can invalidate an
expensive calibration.
2. Industrial Resistance Thermometer Elements
Resistance thermometers could not have migrated from the laboratory to the
industrial plant without the invention of rugged, low-cost elements. The need to
encapsulate the resistive element for protection from shock and corrosive
environments, without unduly straining the element, have led to a variety of
technologies for constructing elements.
Figure diagrams a common ceramic element. The element wire, commonly
platinum, is wound around a ceramic bobbin, welded to lead wires, and coated with glass. The manufacturer must carefully match materials to prevent thermal
expansion strains on the wire.

Higher temperatures are possible with the element construction of Figure. Here,
a coil of platinum wire is passed through bores in a ceramic tube. The bores are
then filled with a ceramic or alumina powder to cushion the coils.

Copper, nickel, and nickel-iron may replace platinum as element materials for
lower temperatures, generally below 260°C. Moderate temperatures also allow
the use of organic materials in element construction, enabling a wide variety of
styles. Resulting benefits include lower cost and faster time response.
Thin-film elements have arrived within the last ten years and are now comparable
to wire-wound ceramic elements in performance, but lower in price.
They consist of a flat substrate with a thin film of deposited platinum, lasertrimmed
to proper resistance. Thin films can have high resistances, commonly
1000 ohms, without the expense of extra platinum wire. Naked elements exhibit
very fast time responses, although they will respond more slowly in probes
because of difficulties transferring heat to the flat element. Also, the user
should be careful to limit measuring current as the small size and high
resistance of thin films invite self-heating.
Thin-films tend to be more fragile than wire-wound elements as well.

3. INDUSTRIAL RTD PROBES
The encased probe is the standard resistance thermometer configuration for
industrial process control and machinery protection. Most probe cases are
stainless steel or Inconel to withstand high temperatures, although other
materials offer advantages at intermediate ranges.

For example, the tip-sensitive probe of Figure has a copper-alloy tip which
conducts heat 20 times better than stainless steel. This design improves
thermal contact with sensed surfaces and reduces errors from conduction along
the sheath.
Standard probe diameters range from 0.125 mm to 0.250 mm.
Smaller probes respond faster when directly immersed, but larger probes may
fit more finely in standard thermo wells.
Probe lengths range from a few inches to ten feet or more.
Figure shows the construction of a high temperature probe. The element fits in
the tip, surrounded by high temperature powder or cement. Extension leads,
normally uninsulated, extend back from the element and are encapsulated by
powder, cement, or bored ceramic spacers. External leads, often insulated with
Teflon or glass braid, are potted with cement at the entry point to seal against
moisture.

The alternative construction of Figure places the element, potting, and lead
transitions within a module at the tip of the probe. This design allows the user
to cut the probe to required lengths. Temperature is limited to the rating of the
external lead wire insulation: 260°C (500°F) for Teflon, up to 550°C (1022°F) for
insulations such as woven mica/glass.

PROBE ASSEMBLIES
A wide variety of mounting fittings and accessories aid probe installation.
Selection depends on the nature of the medium being sensed and cost
requirements.
Direct immersion of a probe into a liquid requires a fitting with a pipe thread, which may be adjustable or welded on the probe. Figure shows a typical
assembly, with one thread for mounting the probe and another for a connection
head. Connection heads provide a transition between probe leads and external
signal wires.
Mounting in a solid material is best accomplished with a spring-loaded holder,
which may be fixed or adjustable.

Spring loading provides good contact of the probe tip against the bottom of the
hole and dampens potentially damaging vibration.

When liquids are particularly corrosive, under high pressure, or fast flowing, a
thermo well may be necessary. A thermo well is a tube, closed at one end,
which protects the probe and allows its removal without breaking the liquid
seal. Many materials and styles are available to match application requirements.
Thermo wells drilled from solid bar stock provide the highest-pressure ratings,
but welded models cost much less. Figure shows a typical thermo well
assembly, including a spring-loaded holder for enhanced thermal response and
reliability.
FLEXIBLE RESISTANCE THERMOMETERS
The encased probes described above do not adapt well to sensing flat surfaces.
Unlike thermocouple junctions, which can be welded directly to metal surfaces,
resistance thermometers present a certain amount of bulk; and heat losses to
ambient air may affect readings.
Small flat elements, such as thin films, may mount on surfaces, but fragile
element and lead wire connections make installation difficult.
Figure shows a flexible resistance thermometer with a wire-wound sensing
element sandwiched between insulating layers. It conforms closely to sensed
surfaces, and has thin insulation to readily transmit heat to the sensing element.
The wire element may be wound to nearly any size to average out temperature gradients, and the flexible construction can withstand extreme shock and
vibration.
Specification of the covering insulation depends on the environment seen by the
sensor.
SPECIAL PURPOSE RESISTANCE THERMOMETERS
Resistance thermometers readily adapt to most process control and thermal
equipment designs. The user may specify cases with axial leads for circuit board
mounting, flat packages for clamping to surfaces, miniature cases for
embedment into metal blocks, and any sheaths and fittings, which can be
produced by a machine shop. In addition, wire windings may be configured to
sense over large areas.
WHERE TO USE RESISTANCE THERMOMETERS
In summary, resistance thermometers offer the greatest benefits relative to
other thermometer types in these situations:
1. Accuracy and stability are the foremost goals of the application
2. Accuracy must extend over a wide temperature range
3. Area, rather than point, sensing improves control
4. A high degree of standardization is desirable.